Calculate and read peaks and integrals

Bind the Dataset to the analysis

Before analysis, Detected peaks (analysis output) is empty and Integration regions (analysis output) prompts you to run NMR Analysis. In Analysis, find NMR Spectrum Analysis, select Run / recalculate now, and choose the Dataset in the primary row. In Input variable bindings, auto uses the canonical variables from the acquired JCAMP. To set the bindings explicitly, map primary.spectral_x to Chemical shift [ppm] and primary.spectral_y to Measured NMR intensity.

Analysis settings for the ¹H region from −2 to 6 ppm.
Analysis settings for the ¹H region from −2 to 6 ppm.

Set the range, baseline, and resolution

GUI controlDefault valueHow to use it
Lower chemical shift [ppm] and Upper chemical shift [ppm]−5 and 220Set the lower and upper limits of the range to process. Check that the trace contains samples in that range.
Baselinelinearnone keeps the baseline. constant subtracts a constant estimated from the median. linear subtracts an estimated linear baseline. Which method to use depends on the shape of the background in the analysis range.
Peak prominence [normalized]0.05Threshold for how much a peak stands out from its local background after normalization by the largest absolute baseline-corrected value. Lower thresholds admit weaker signals and can also admit noise.
Minimum peak separation [ppm]0.15Minimum separation between detected peaks. A smaller value can retain neighboring maxima if the samples resolve them.
Peak search samples (extrema preserving)8192Number of samples used for peak detection only. When the search grid is reduced, local extrema and their coordinates are retained. Peak widths estimated on a reduced grid are approximate.
Integration lower bound [ppm] and Integration upper bound [ppm]0 and 220Subrange of the processed region to integrate. It must contain at least two samples. The integral uses the trapezoidal rule on normalized intensities and can be negative when a negative lobe dominates the window.
Maximum analysis samples (full-resolution default)131072Limits the effective resolution of the entire analysis. If the limit is below the region’s sample count, the quantitative calculation also uses the reduced grid. Leave it at 131072 for the exercises.

Run the analysis and read the tables

In Run, check Execution plan and click Run enabled analyses. Wait for the completion message. In Dataset, the calculation publishes .nmr.processed, .nmr.display, .nmr.peaks, and .nmr.integrals. The acquired Dataset remains separate. Return to NMR Spectrum and select the primary again. Read Shift [ppm] for the peak position, Intensity for normalized intensity, and Prominence to see how much the peak stands out from the local background. The table displays rounded values. Scroll within it to see additional rows. Integration regions shows the limits in ppm and the integral value.

The numerical detector reports peaks. Assigning them to chemical groups requires information about the sample, calibration, phase, experimental conditions, and suitable spectral references.

Selecting the analyzed Dataset shows the peaks and integral for the requested region.
Selecting the analyzed Dataset shows the peaks and integral for the requested region.