Select the source, overlay, and plot

Select the spectrum and check the metadata

In NMR source, choose the Dataset. Check Nucleus, Frequency [MHz], Solvent, and Reference below the selection. They show the observed nucleus, frequency, solvent, and chemical shift reference metadata. Compare them before overlaying two traces. Source type gives the original file’s domain, Processing describes how the trace was prepared, and Source SHA-256 gives the hash of the original. Acquiring the same JCAMP file with different settings can produce distinct Datasets with the same source hash.

The ¹H source shows NMR FID as its original type and fid_fft as the preparation for the ppm plot.
The ¹H source shows NMR FID as its original type and fid_fft as the preparation for the ppm plot.

Overlay two compatible traces

In Compatible overlay, leave the selection blank to show only the primary spectrum. To overlay another spectrum, choose its ID. The list shows other Datasets for the same nucleus. Switching the primary to ¹³C removes an incompatible ¹H overlay. The list filters by nucleus. Check frequency, provenance, chemical shift reference, phase, range, and intensity scale separately.

Before comparing two processing runs on the same FID, you can run the same analysis over the same ranges. The relative scale of the processed traces depends on each Dataset’s normalization. Any Detected peaks markers belong to the primary spectrum.

Choose the ppm axis direction

In ppm direction, reverse puts higher values on the left, following the NMR convention. normal shows increasing values to the right. Changing direction affects the figure while preserving the ppm coordinates and numerical results. Create a new variant with Open persistent NMR graph to see the change.

Create and reopen the persistent plot variant

  1. Choose the source, overlay, and direction. Click Open persistent NMR graph. GX-TXT creates a plot.dataset.xy/pNNNN variant and opens PlotSpec v2 Series. The title shows the variant number.
  2. In the editor, check the series using Dataset, X, Y, Kind, and Label. A single trace shows spectral_x, spectral_y, and line. After analysis, scatter may appear for peaks in the primary spectrum. Traces can point to .nmr.display if the analysis published it. The plot uses that display series. The integral was calculated on the analysis Dataset.
  3. To view the figure, open Plots in the main window. Use Refresh if needed, select Plot pNNNN, and open Graph → Open PNG 1:1. In the image window, use Fit to fit both axes. Zoom + and Zoom – change the zoom. Reset returns to 1:1. Drag with the left mouse button to draw a rectangle, then use Fit area to fit the selection. Without a rectangle, the viewer displays Select an area first. Show annotations toggles any annotations on the figure. The variant belongs to the session and remains accessible after closing the NMR viewer. Reopening the viewer from its tile in the same session restores the source, overlay, and direction used for the plot.
The editor lists the display trace and peak points for the NMR plot variant.
The editor lists the display trace and peak points for the NMR plot variant.

Reopen the viewer after switching sessions

Reopen Viewers → NMR Spectrum after each session switch. In the tested version, a viewer left open can continue to show sources and overlays from the previous session. Trying to plot in an empty session gives Dataset … is not an acquired NMR spectrum. Reopening the viewer from its tile shows the correct state and disables the plot button. In a session with spectra, check the ID and metadata in the new window before plotting.

After reopening the viewer from its tile in an empty session, the previous Datasets are absent and the button remains disabled.
After reopening the viewer from its tile in an empty session, the previous Datasets are absent and the button remains disabled.