Five exercises with real spectra

1. Open the previously processed ¹H spectrum

  1. Import cambridge_1h_gia_processato.dx as yp80f-1h-misurato with the default FID settings of 4, 2 Hz, and 0°. Reopen the viewer and select the ID.
  2. Check NMR SPECTRUM and already_processed, with an observation frequency of about 400.33 MHz. Leave the overlay blank and keep the direction set to reverse. Open the persistent plot, then open the figure from Plots.

The plot shows one line. Its 131,072 points span approximately −25.02 to 24.94 ppm. Peaks and integrals are published after analysis. The main signal is near 2.29 ppm in the calibrated original.

Trace acquired from an already processed JCAMP file. The plot shows the NMR axis direction and one series.
Trace acquired from an already processed JCAMP file. The plot shows the NMR axis direction and one series.

2. Process a ¹H FID and read the peaks and integral

Import cambridge_1h_fid.dx as yp80f-1h-fid with settings of 1, 2 Hz, and 0°. The metadata show NMR FID and fid_fft. In NMR Spectrum Analysis, set the range to −2 to 6 ppm, the baseline to linear, prominence to 0.03, separation to 0.06 ppm, peak search samples to 4096, integration to 0.2–1.2 ppm, and the analysis limit to 131072. Run the analysis and select the source again.

The calculation found 11 peaks. The peak at 0.6209 ppm has a normalized intensity of 1. The integral from 0.2 to 1.2 ppm is approximately −0.104588. Open the plot with reverse and leave the overlay blank. The peak series marks the maxima selected by the detector. The negative lobe shows why the integral has a sign.

Negative intensities remain visible. Markers distinguish the detected peaks from the processed trace.
Negative intensities remain visible. Markers distinguish the detected peaks from the processed trace.

3. Compare two processing runs on the same FID

Acquire cambridge_1h_fid.dx again as yp80f-1h-variante with settings of 4, 4 Hz, and 25°. Run the same analysis as in exercise 2 on the second Dataset. The maximum is at 0.6621 ppm and the integral over the same range is about −0.126952. Both Datasets retain the same Source SHA-256.

In the viewer, choose yp80f-1h-fid as the primary, yp80f-1h-variante as the overlay, and reverse as the direction. Open a new persistent plot. The legend lists two ¹H series and Detected peaks for the primary. Changing three settings together shows their combined effect. To isolate one contribution, vary one parameter per acquisition.

Two processing variants of the same measurement share the ppm axis and use different phase and window settings.
Two processing variants of the same measurement share the ppm axis and use different phase and window settings.

4. Analyze the carbon-13 spectrum and change the axis direction

Import cambridge_13c_solido_fid.dx as yp80f-13c with settings of 1, 2 Hz, and 0°. Select it as the primary and check for ^13C and about 100.67 MHz. The ¹H spectra disappear from the overlay list. Set the analysis range to 75–210 ppm, the baseline to constant, prominence to 0.03, separation to 0.2 ppm, peak search samples to 4096, integration to 105–139 ppm, and the analysis limit to 131072.

The detector reported seven peaks. The most intense is at 119.7609 ppm and the integral is about 0.003442. Set normal and open the variant. The ppm labels increase to the right. The axis follows the limits of the analyzed range.

The ¹³C plot shows increasing ppm values and markers for the seven peaks.
The ¹³C plot shows increasing ppm values and markers for the seven peaks.

5. Explore a solution-state ¹³C spectrum and filter the peak results

Import cambridge_13c_soluzione_fid.dx as titolazione-13c with settings of 1, 2 Hz, and 0°. The viewer shows about 150.90 MHz and CD3CN as the solvent. Set the analysis range to 95–195 ppm, the baseline to none, prominence to 0.025, separation to 0.1 ppm, peak search samples to 8192, integration to 145–174 ppm, and the analysis limit to 131072.

The detector found the maximum at 160.1427 ppm with an intensity of 1. The integral is about −0.003555. The low threshold returned 515 rows. Scroll within the table to review them. In this representation, the smaller maxima can include signal components, artifacts, or noise. Raising the prominence threshold limits the results to more distinct maxima.

The titration example returns to reverse axis direction. The low threshold produces many markers in weak regions.
The titration example returns to reverse axis direction. The low threshold produces many markers in weak regions.