Fractional-Octave Band Analysis

GX-TXT DEV112 P011-bis

Group the spectral components into fractional-octave bands and read levels and center frequencies in the result.

Enable Fractional-Octave Band Analysis

Open Configuration → Modules → Analysis & Services and select Fractional-Octave Band Analysis (analysis.fractional_octave_spectrum, 1.2.0 in P011-bis). Check Health Available and press Enable module when it is off. In Analysis check Run / recalculate now before submitting the computation.

The module is available and can be enabled from the Analysis & Services catalog.
The module is available and can be enabled from the Analysis & Services catalog.

Generate the test spectral density

With New Session… → Generate Session… open GX-TXT Public Session Generators → Signal Processing → Fractional-Octave PSD, fractional_octave_psd 1.0.1. Choose Standalone Session, give it a name and generate. The lab provides fractional_octave.source, a synthetic linear PSD with a 1 Hz step from 0 to 4096 Hz. The bins of the tones at 125, 500 and 2000 Hz contain 0.25, 0.09 and 0.04 V² respectively, plus a uniform floor. In Dataset → Preview identifies the source. From Technical details check the contract spectrum.table and the units Hz and V^2/Hz.

The selected Generator publishes a synthetic linear PSD with three known tones.
The selected Generator publishes a synthetic linear PSD with three known tones.

Choose frequency, density and bands

In Dataset sources assign fractional_octave.source to primary. Its Remove is disabled. The bindings Frequency variable and Linear PSD (power per Hz) must match primary.frequency_Hz and primary.psd_V2_per_Hz. The menu also offers auto. Check that the abscissa in Hz increases and that the density is linear and non-negative. For an output of the Welch module use psd, the power density.

Bands per octave
1 uses whole octaves, 3, the default, uses third-octaves and 6 uses sixths. The computation uses exact base-10 bands with an octave ratio of 103/10 and a reference center of 1000 Hz. More bands increase the detail.
Frequency range
auto uses the acquired support. explicit applies Minimum frequency [Hz] and Maximum frequency [Hz], initially 20 and 20000. Use positive, ordered limits. The complete bands compatible with the data support and the chosen interval are published. The 6-band variant with 100–5000 Hz returns 31 complete bands in the lab, whose support ends at 4096.5 Hz.
RMS reference for band level
Positive RMS amplitude in the same unit as the signal. It starts from 1 V in the test. The level is 20 log10(band RMS/reference RMS), in dB. Changing the reference changes the level, while power and RMS stay the same. In the third-octave test, 0.1 V raised the levels by 20 dB with the same power and RMS.
Reset
Resets the single numeric field to its initial value.
The six-bands-per-octave variant uses an explicit frequency interval.
The six-bands-per-octave variant uses an explicit frequency interval.

Read integrated power, RMS and level

Open Run, check the module and start. After completion open the Dataset fractional_octave.source.frac_oct.v0002.n3 for the third-octaves. The variants use .n1 and .n6. In Preview read the identity, in Open Dataset → Canonical table the values and in Metadata and Technical details the origin, the HDF5 format and the structure.

Column
band_number, center_frequency
Reading
Band index and center. The exact center frequency can differ from the rounded nominal frequency.
Column
lower_frequency, upper_frequency, bandwidth
Reading
Band limits and width in Hz.
Column
band_power, band_rms
Reading
Integral of the density over the band in V² and its square root in volts.
Column
band_level_db, coverage_fraction
Reading
Level referred to the chosen RMS, and share of the band covered by the support. In the complete product the coverage is 1.

In Results open fractional_octave_report.txt. In the third-octave test 35 bands are published. The center at about 125.893 Hz has the maximum power, about 0.25003 V². The sum of the complete bands is about 0.38355 V². The procedure integrates a numerical PSD, so an instrument conformance check requires its own specifications and tests.

Canonical table of the computed bands and their units.
Canonical table of the computed bands and their units.

Open and customize the band plots

In Plots, 01_band_level_db shows the level in dB, 02_band_rms the RMS voltage and 03_band_power the power. Frequency is the horizontal coordinate. Read axes and limits to interpret the amplitude. In the variant Plot p0001 you can change the title with Presentation → Title and labels…, save, choose Render → Rerender and open the new PNG from Graph → Open PNG 1:1. The change concerns the Session plot.

Band level variant before customization.
Band level variant before customization.
Band plot reprocessed with the title set in the Plots panel.
Band plot reprocessed with the title set in the Plots panel.