Quantitation with UV/Vis Spectral Families
GX-TXT DEV112 P011-bis
Configure UV/Vis quantitation on a family of spectra and check the outputs of the example with known concentrations.
GX-TXT P011-bis · module 1.2.0. UV/Vis Spectral Family Quantitation extracts the absorbance at one wavelength, fits a linear calibration in concentration and provides responses, residuals, validation and, if requested, the estimate for an unknown.
Prepare the Manchester standards
From Session → New Session → Generate Session choose Manchester MXene UV/Vis and Standalone Session. The measured spectra were published by James Bird under the CC BY 4.0 licence, DOI 10.48420/21354474.v1. Manchester data and lab notebook.
Follow Python Script to publish the families with the canonical schema. The Dataset uvvis.manchester.concentration keeps ten members and 8,010 rows. In Datasets → Preview and Metadata check member_id, parameter_value in wt%, spectral_x in nm and spectral_y in absorbance. A CSV family with column names only and different stable IDs was rejected. Use the schema with the four canonical IDs.

Configure calibration and validation
In Configuration → Modules → Analysis & Services select UV/Vis Spectral Family Quantitation, check Health and choose Enable module. In Analysis tick Run / recalculate now.
- Dataset sources
- Assign
uvvis.manchester.concentrationtoprimary. Assignuvvis.blanktoblankonly for the subtraction branch, and a spectrum to estimate tounknown. Leave the unknown source empty when you only want the calibration. The rows are managed with + Add source and Remove. The test usesuvvis.sample.9as the teaching unknown, the same as the member kept out of the fit in the first Run. - Analytical wavelength [nm]
- Queries each spectrum at the same wavelength within the covered domain. The default is 280 nm. The first pass uses 322 nm. The second uses 562 nm.
- Held-out validation member IDs (semicolon separated)
- Lists the IDs of the members to exclude from the fit and use for checking. Separate them with
;. The first pass uses9, the second leaves it empty and fits all ten standards. - Absorbance blank correction
noneuses the available absorbances.subtract_blank_absorbancesubtracts the blank at the corresponding wavelengths and requires the sourceblank. Use the correction if it is consistent with the preparation, and check the coverage of the blank. The blank variant here is a teaching comparison.- Peak search minimum / maximum [nm]
- Define the search interval for the suggested spectral maximum, initially 250–800 nm. The suggestion is distinct from the analytical wavelength you chose for the fit.

Start and interpret the calibration
Open Run, update the plan and start the selective analysis. In Results read uvvis_report.md. The test with blank subtraction at 562 nm and ten standards gave a slope of about 224.329 absorbance units per wt%, an intercept of 0.033758 and R² 0.992925. The suggested maximum is 320 nm and the range of the standards is about 0.0000608–0.001276 wt%. The residual plot helps check linearity, and the estimate must be interpreted against the calibrated range.
.calibrationkeeps the calibration quantities. In Canonical table check values and units..selected_responsescollects responses and members..lambda_maxdocuments the suggested maximum..residualsshows the differences between responses and fitted values..validationevaluates the held-out members, when requested..unknown_resultrecords the estimate from the sourceunknownwhen provided..blank_corrected_familyis published with blank subtraction active. In Metadata and Technical details read the provenance of the products.


Check overlay, fit and residuals
Open Plots and select the family overlay, the calibration fit or the residuals. Use the overlay to choose an interpretable wavelength and the residual plot to look for persistent structure. In an editable variant of type Plot p…, Presentation → Title and labels… lets you change title and axes. In the column Use leave ticked the variants to update. Use Render → Rerender selected and open Graph → Open PNG 1:1.

If in the same Session you generated a calibration with validation and then cleared Held-out validation member IDs, a previous variant can keep the series .validation/response. In the tested build the numerical Run finishes, while the plot regeneration can stop with no finite rows in the selected X range. By setting member 9 again and repeating the Run, the test republished the series. With the current calibration and response variants ticked in Use , Rerender selected then completed the regeneration. Use the table Canonical table and the report to look at the case without validation. Keep the results apart before recalculating with different members.